Topcon Bm-7a And Sr-3ar: The Ultimate Setup Guide

Published:

Updated:

Disclaimer

As an affiliate, we may earn a commission from qualifying purchases. We get commissions for purchases made through links on this website from Amazon and other third parties.

Setting up display calibration devices such as the TopCon BM-7A colorimeter and SR-3AR spectroradiometer requires specific configurations and connections. This article serves as a comprehensive guide for the ultimate setup of these devices. The BM-7A colorimeter necessitates the use of a straight-through 9-pin female-to-female serial cable, while the SR-3AR spectroradiometer requires a null modem 9-pin female-to-female serial cable. Additionally, the BM-7A colorimeter requires the adjustment of configuration switches and connection to a serial port on the Calman computer or a Serial to USB adapter. Similarly, the SR-3AR spectroradiometer needs specific menu options to be set and should be connected in a similar manner. Different meter modes and standard exposure modes should be selected for each device. Furthermore, the SR-3AR spectroradiometer can provide full spectral data if the appropriate option is checked in the Calman settings. This guide ensures a meticulous and precise setup process for the TopCon BM-7A and SR-3AR devices, catering to an audience that seeks innovative solutions.

Key Takeaways

  • Calman version 5.7.1 or later is required for TopCon BM-7A and SR-3AR setup.
  • A straight-through 9-pin female-to-female serial cable is used for BM-7A colorimeter connection.
  • A null modem 9-pin female-to-female serial cable is used for SR-3AR spectroradiometer connection.
  • The configuration switches and settings for BM-7A and SR-3AR differ, including baud rate, length, parity, stop bits, and delimiter.

Required Version

Version 5.7.1 or later is necessary for the setup of TopCon BM-7A and SR-3AR, as stated in the pre-existing knowledge. This version of the software provides the required compatibility and functionality to ensure a successful setup of the TopCon devices. It is crucial to use the specified version to avoid any compatibility issues and to take advantage of the latest features and improvements offered by the software. By using the recommended version, users can ensure a seamless integration of the BM-7A colorimeter and SR-3AR spectroradiometer with Calman. This compatibility ensures accurate and reliable measurements, allowing users to achieve precise color calibration and achieve optimal display performance. Keeping the software up to date is essential for users who seek innovation and want to leverage the full potential of their TopCon devices.

BM-7A Colorimeter Connection

The BM-7A colorimeter connection requires the use of a specific serial cable and certain configuration switches to be set in a particular manner. To establish the connection, a straight-through 9-pin female-to-female serial cable is needed. It is important to note that the cable used for the BM-7A colorimeter connection is different from the one used for the SR-3AR spectroradiometer. The configuration switches on the BM-7A colorimeter should be set as follows: Switches 1, 2, and 5 are down, while switches 3, 4, 6, 7, and 8 are up. The connection is made by connecting the BM-7A colorimeter to a serial port on the Calman computer or to a Serial to USB adapter using the straight-through serial cable. On the Calman Meter Settings tab, the user should click on Find Meter and select the appropriate Com Port.

SR-3AR Spectroradiometer Connection

To establish a connection with the SR-3AR spectroradiometer, a null modem 9-pin female-to-female serial cable is required, which is different from the cable used for the BM-7A colorimeter connection. The SR-3AR spectroradiometer should be set to a baud rate of 38,400, with a length of 7, parity set to even, stop bits set to 1, and delimiter set to CR+LF. It should be connected to a serial port on the Calman computer or to a Serial to USB adapter using the null modem serial cable. On the Calman Meter Settings tab, the Find Meter button should be clicked, and in the Find Meters popup dialog, the Com Port should be set to auto or the assigned com port for the meter or Serial to USB adapter. This setup ensures a proper connection between the SR-3AR spectroradiometer and Calman for accurate and stable measurements.

BM-7A Colorimeter Connection Procedure

The connection procedure for the BM-7A colorimeter involves setting the configuration switches on the device, adjusting the baud rate, length, parity, stop bits, and delimiter settings, and connecting it to a serial port on the Calman computer or a Serial to USB adapter using a specific type of serial cable. To ensure a successful connection, follow these steps:

  1. Set the configuration switches on the BM-7A colorimeter as follows: Switches 1, 2, and 5 are down. Switches 3, 4, 6, 7, and 8 are up.
  2. Configure the baud rate, length, parity, stop bits, and delimiter settings on the colorimeter. Set the values as follows: Baud rate = 38,400; Length = 7; Parity = Even; Stop Bits = 1; Delimiter = CR+LF.
  3. Connect the BM-7A colorimeter to a serial port on the Calman computer or to a Serial to USB adapter using a straight-through serial cable.

By following these steps, you will be able to establish a reliable connection between the BM-7A colorimeter and the Calman software, allowing for accurate and precise measurements in your color calibration process.

SR-3AR Spectroradiometer Connection Procedure

For successful integration of the SR-3AR spectroradiometer into your calibration process, a specific connection procedure needs to be followed. To begin, set the SR-3AR spectroradiometer menu options as follows: Baud rate = 38,400; Length = 7; Parity = Even; Stop Bits = 1; CS-900 On/Off = Normal Type; Delimiter = CR+LF. Next, connect the SR-3AR spectroradiometer to a serial port on the Calman computer or to a Serial to USB adapter using a null modem serial cable. Then, on the Calman Meter Settings tab, click Find Meter. In the Find Meters popup dialog, leave the Com Port set to auto or select the assigned com port for the meter or Serial to USB adapter. Following these steps will ensure a successful connection of the SR-3AR spectroradiometer for your calibration needs.

Meter Mode for BM-7A Colorimeter

The Meter Mode for the BM-7A Colorimeter is an essential setting that determines the correction factor used for measuring the display. It allows users to select either the factory correction factor or a user-defined correction factor that corresponds to the display being measured. This setting ensures accurate and precise measurements by adjusting for any inconsistencies or variations in the colorimeter’s performance.

When configuring the Meter Mode for the BM-7A Colorimeter, users should consider the following:

  • Select the appropriate correction factor based on the target display type.
  • Choose a correction factor that yields stable measurements in Calman Read Continuous mode.
  • Ensure that the selected correction factor aligns with the specific characteristics of the display being measured.

By carefully adjusting the Meter Mode, users can optimize the performance of the BM-7A Colorimeter and obtain reliable and accurate measurements of their displays.

Meter Mode for SR-3AR Spectroradiometer

When configuring the Meter Mode for the SR-3AR Spectroradiometer, users have the option to select from a range of available correction factors to ensure accurate measurement of various display types. Under the Meter Mode (Target Display Type) setting, users can select "All Types" to cover a wide range of displays. This allows the SR-3AR Spectroradiometer to provide precise and reliable measurements for different types of displays. By selecting the appropriate correction factor, users can account for any variations in the display’s characteristics and ensure accurate color measurement. This flexibility in Meter Mode enables users to obtain accurate and consistent results when using the SR-3AR Spectroradiometer in conjunction with Calman software. Whether it’s for professional color grading, quality control, or research purposes, the Meter Mode options provide the necessary tools for innovation and advancement in display measurement technology.

Full Spectral Data with SR-3AR Spectroradiometer

When using the SR-3AR spectroradiometer with Calman, it is possible to obtain full spectral data on every reading. This feature can be enabled in the Calman settings under "Application Measurement Options" by checking the "Use Software Color Matching Functions" option. By doing so, the SR-3AR spectroradiometer will provide a comprehensive analysis of the entire spectrum, allowing for more accurate color measurements and analysis. This capability is particularly useful for users who require detailed spectral data for their research or calibration purposes. With the ability to capture and analyze the full spectrum of light, the SR-3AR spectroradiometer offers a high level of precision and versatility in color measurement applications.

Advantages of Full Spectral Data with SR-3AR Spectroradiometer:

  • Enhanced accuracy in color measurement and analysis
  • Comprehensive analysis of the entire spectrum for detailed research and calibration purposes
  • Versatility in various color measurement applications.

About the author

Latest posts

  • Maintaining Consistent Color Accuracy Across Various Displays In A Single System

    Maintaining consistent color accuracy across various displays in a single system is crucial for ensuring an optimal visual representation. In today’s digital age, where images and videos are widely used for communication, accurate color reproduction plays a significant role in conveying the intended message effectively. However, achieving this consistency poses several challenges due to variations…

    Read more

  • Maintaining Consistency In User Experience Across Multiple Projectors For Better Image Quality

    Maintaining consistency in user experience across multiple projectors is crucial for achieving optimal image quality. This article explores the importance of uniform calibration, aligning projection angles and positions, utilizing advanced blending techniques, managing color and brightness consistency, as well as regular maintenance and testing. Uniform calibration ensures that all projectors in a multi-projector setup display…

    Read more

  • Lens Shift Vs Zoom: Which Setting Should You Prioritize For Your Business Needs

    In today’s rapidly evolving business landscape, technological advancements have become a crucial component for success. When it comes to photography and visual communication, professionals are faced with the decision of prioritizing either lens shift or zoom settings for their specific business needs. Lens shift refers to the ability to adjust the position of the lens…

    Read more